LS 6000 Scintillation System User’s Manual
PN 247971-F
5-22
Data Calculation
Single Photon Monitoring
Count Time per Data Point
This feature allows you to enter the amount of time to accumulate counts for each data point.
This time is restricted by the Counting Time/Sample and Number of Data Points. The
mini-mum count time is 0.05 minutes (3 seconds). The maximum count time is the Counting
Time/Sample divided by the Number of Data Points. The Number of Data Points X the Count
Time per Data Point cannot exceed Sample Counting Time.
To change this setting, highlight “Count Time/Data Point”, and enter the counting time.
Count Sample Set
The entire set of samples can be recounted a number of times. When another User Number
Card or Halt Card is detected, the sample changer backs up to the beginning of the sample set
and recounts the samples. This is repeated until the entry for Count Sample Set is satisfied.
To change the number of times to count the sample set, highlight “Count Sample Set” and
enter the desired number.
Factor
This feature enables you to obtain a printout in which the count data have been multiplied by
a constant:
CPM X Factor
= Final Answer Printed Out
To change the factor, highlight “Factor” and enter a new factor. The factor may have a
maximum of nine digits and decimal point. Values greater than or less than this range may be
entered as an exponential using a 10 character field. For example, the number 1.0623 X 10
-23
is entered as 1 O623E
-23
Loading Samples for Single Photon
Samples are loaded into the racks in the order you wish to count them. The Single Photon
Program does not recognize blanks or replicates.
Results for Single Photon
To count the samples, refer to Section 3.2 and Section 3.3. An adapted counting routine is
present in the Count Single Rack mode, when a User Program with SPM is selected, which
Vista de pagina 99