LS 6000 Scintillation System User’s Manual
PN 247971-F
6-21
Isotope/DPM Libraries
Setting Up A Quench Curve
6
7. Highlight “Add or Delete Quench/Efficiency Pairs” and press SELECT. The Main Editing
Window displays the menu shown in Figure 6.16.
8.
To add points to the quench curve, type in the H# or IC# value and the Efficiency for each point
you wish to add. Press ENTER after each value is typed in. Use the Cursor Arrow keys to move
to the next entry. A maximum of ten points for each quench curve may be stored.
9. 9. To delete points from the quench curve, highlight any value of the point you wish
to delete, and press DELETE. Delete other values in the same manner. At least five
points must remain after all desired points have been deleted.
10. A prompt at the bottom of the menu allows you to specify whether the data was
generated using standards that are colored. If the standards are colored, highlight the
prompt and choose Yes.
11. The background quench curve may also be edited. Highlight “Color Quench Stds” and
press the Down Cursor Arrow key. The menu shown in Figure 6.17 is displayed. You
may choose to use the background curve or not. Highlight “Use Background Curve” and
choose the desired setting. To change or input a background quench curve, highlight the
Figure 6.16 Menu To Add/Delete Points From Quench Curve.