LS 6000 Scintillation System User’s Manual
PN 247971-F
b-1
Installation Requirements
Electrical Requirements
b
b
Installation Requirements
This appendix describes the electrical power, space and environmental conditions required
for installation of a Beckman Liquid Scintillation instrument. Proper preparation of the
installation site will minimize installation time and allow optimum performance from your
Beckman LS System.
NOTE For proper warranty validation, all Beckman Liquid Scintillation Systems must be installed by an
Authorized Beckman Field Service Representative. Installation is included in the purchase price of the
instrument. When the instrument arrives at your facility, contact your local Beckman Sales and Service Office
for an installation appointment.
b.1 Electrical Requirements
Beckman Liquid Scintillation Systems require a minimum of two power outlets; one for the
basic counter and one for the printer. All power cords are 2.2m (86 inches) long.
For proper function and reliable
test results from any laboratory instrument, certain power
line requirements must be satisfied. the following are excerpts from the NCCLS Power Source
Standard 5, POWER
REQUIREMENTS FOR LABORATORY INSTRUMENTS.
1
Power Supply
Power Lines that supply clinical laboratory instruments should be reserved exclusively for
service within the clinical laboratory. Each power line to a clinical laboratory should connect
directly from a main power line transformer at a power source known to be clear of erratic
power loads, spikes, and electromagnetic interference.
Electrical Requirements:
(50/60 Hz) 120V 240V BTU/HR
Instrument 3A 1.5A 1230
Temperature 7.5A 3.8A 3060
Control Accy
Inrush Current: 5A at 120V
2.5A at 240V
1Power Requirements for Clinical Laboratory Instruments and Laboratory Power Lines,”
NCCLS TENTATIVE STANDARD: TSI-5.