LS 6000 Scintillation System User’s Manual
PN 247971-F
3-19
Counting Samples
Multi-Task
3
Isotope Library Section 6 for adding and deleting isotopes if you are not familiar with
these tasks.
NOTE You may not set up an isotope in Multi-Task using Automatic Setup.
5. When you are finished using New Isotope Setup, highlight “Isotope Selection Complete’
and press SELECT. The Multi-Task Menu is displayed.
6. Press PREVIOUS MENU to return to the Auto Count display.
Using the DPM Library in Multi-Task
The DPM Library can be accessed during an Automatic Count using Multi-Task. Stored
quench curves can be reviewed, edited or deleted. A quench curve not currently stored in the
DPM Library may be set up manually.
To set up a new quench curve:
1. During Automatic Count, press SELECT to display the Multi-Task Menu shown in
Figure 3.8.
Figure 3.17 Select Another Isotope Menu in Multi-Task.