LS 6000 Scintillation System User’s Manual
PN 247971-F
3-12
Counting Samples
Multi-Task
NOTE You may abort the Interrupt Count and resume Automatic Count by pressing [SELECT] at any time
after this point.
7. When the samples in the Interrupt Rack have been counted, the instrument resumes the
previous Automatic Count where it was interrupted.
Results of Interrupt Count are displayed and stored. They are not printed out. Refer to
Interrupt Data Section 3 for information on accessing Interrupt Data.
Interrupt Count Using a User Program During Multi-Task
To count with a User Program:
1. Highlight “Select User Program” from the Count Single Rack Menu (Figure 3.3) and
press SELECT. The User Program Selection Menu (Figure 3.5) is displayed.
2. 2. Select the desired User Program. The Main Editing Window displays a summary of
the program. A typical summary is shown in Figure 3.10.
3. To change only the counting time, highlight “Counting Time” and enter the desired
count time. Highlight “Count With Pro-gram of User:” and press SELECT. Refer back to
step 5 above to continue Interrupt Count.
4. To change other counting parameters, highlight “Change Counting Parameters” and
press SELECT. The Main Editing Window displays the parameters of the User Program.
Figure 3.10 Summary of User Program in Interrupt Count