LS 6000 Scintillation System User’s Manual
PN 247971-F
2-18
Getting Started
Calibration
To change these parameters, highlight “AUTO DPM Calibration Setup” and press SELECT.
The menu shown in Figure 2.10 is displayed. Highlight each prompt and enter the values for
the appropriate isotope.
Color Selection - Monitor
If a color monitor is installed, this parameter is used to select the colors displayed on the
monitor and the colors displayed on the graphics window. To change the settings, highlight
“Color Selection” and choose the desired color for both the normal display and the graphic
display.
2.7 Calibration
Calibrating is done to ensure that when an isotope is chosen, the window setting covers the
energy spectrum for the isotope.
Calibrating the Instrument
Calibration is performed using the Calibrate Rack set up as described in Section 2 Setting Up
the Racks. Calibration can be performed before beginning an Automatic Count or whenever
desired. Refer to Section 2 When Calibration Should Be Done for recommendations.
To calibrate:
1. Place the Calibrate Rack on the right-hand side of the sample changer, in the rear, so it Is
the first rack counted.