
LS 6000 Scintillation System User’s Manual
PN 247971-F
6-22
Isotope/DPM Libraries
Setting Up A Quench Curve
appropriate prompts and enter the values. To display the previous menu, highlight the
first prompt and press the Up Cursor Arrow key.
12. Once all points are added or deleted and the background quench curve Is edited, press
START to print the new correlation table. Figure 6.13 shows a typical correlation table
for a SL
3
H quench curve.
13. Press PREVIOUSMENU] to store the edited quench curve. After storing the edited
curve, the menu shown in Figure 6.14 Is displayed.
If you do not want to store the changes after editing, press CANCEL. The changes are
not stored.
14. HighlIght “Review/Edit Complete” when you are finished editing and press SELECT.
Press MAIN MENU if you are finished using the DPM Library.
Deleting A Quench Curve
If a quench curve Is no longer required or if a new curve Is to be generated for a particular
Isotope, the existing curve may be deleted from the library. Deleting a quench curve affects all
User Programs using that quench curve.
To delete a quench curve:
Figure 6.17 Menu to Edit Background Quench Curve
Comentarios a estos manuales