LS 6000 Scintillation System User’s Manual
PN 247971-F
5-15
Data Calculation
DPM % of Reference
5
NOTE Do not leave an empty space for each missing replicate. One empty space indicates to the system that
a new group of replicates follows.
Results for Xtal DPM
To count the samples, refer to Section 3.2. A typical printout in standard format for Xtalscint
DPM is shown in Figure 5.13.
5.7 DPM % of Reference
The DPM % of Reference program is used for any type of experiment that requires dMding
each of the unknowns by a reference, yielding a percent value. The data is first corrected to
dpm using the quench curves stored in the DPM Library. Refer to Section 5.5 for more
information on DPM. Blanks and backgrounds can be subtracted if selected in the User
Program.
Figure 5.13 Typical Printout for Xtal DPM
% Ref
Sample DPM - Blank CDM x 100 x Factor
Reference DPM - Blank DPM
=