LS 6000 Scintillation System User’s Manual
PN 247971-F
c-15
RS232
RS232 Input Mode
c
stopped, but counting will
continue for
the remaining samples in this set.
998 Process Fatal Message Integer
The current function must be
stopped.
In Automatic Counting, the
current user
is aborted, but the count will
continue with the next user, if any.
999 Instrument Fatal Message Integer
A hardware failure prevents
further
instrument operation.
This Includes power failures when
the
current function cannot be
resumed.
Figure c.5 Datum Ids.
Figure c.6 Datum ID’s for Alpha-Beta Discrimination.
Setup Parameters
Data Id Values: 100 - 199
ID Data Value Meaning Data Value Type
164 Alpha-Beta Calibration Name Text (20 Char)
165 Alpha-Beta Selection On/Off Text (3 Char)
166 Left Rvalue Limits Text (15 Char)
(2 Reals Separated by the Following 3
Character String: ‘-‘)
167 Right Rvalue Limits Text (15 Char)
(2 Reals Separated by the Following 3
Character String: ‘-‘)
168 Pulse Height Limits Text (15 Char)
(2 Reals Separated by the Following 3
Character String: ‘-‘)
169 Beta Half-Life Correction Date Text (17 Char)