LS 6000 Scintillation System User’s Manual
PN 247971-F
6-17
Isotope/DPM Libraries
Setting Up A Quench Curve
6
in one rack and contain at least five and no more than 10 quenched standards of the
particular isotope. The vials in a set need not be in order of quench.
17. When the racks are loaded into the sample changer, press START. The samples are
counted to 0.5% 2 sigma or for 20 minutes each.
Figure 6.13 Typical Printout of Correlation Table.