LS 6000 Scintillation System User’s Manual
PN 247971-F
2-2
Getting Started
Powering Up
The only user accessible fuses are the power fuses, located next to the receptacle where the
power cord is plugged into the instrument. The instrument has two fuses.
Parts required:
100/ 120V instrument—4A time lag fuse (UL/CSA),P/N 801579
220/240V instrument—2A time tag fuse (lEO), P/N 898227
Tools required: Small flatblade screwdriver.
1. Locate the power cord receptacle, located on the left-hand side of the instrument, in the
lower rear corner. Try not to move the instrument. If it is necessary to move the
instrument, move it carefully, so that the lead blocks do not harm the photomul-tiplier
tubes.
2. Unplug the power cord from the instrument.
3. Use the screwdriver to remove fuses from the fuse holder.
Figure 2.2 Tools Required
110/120 v INST 220/240 V INST
Two 801579 Two 898222
4 A Time Lag Fuses 2 A Time Lag Fuses
0.25” X 1.25” (5 mm X 20 mm)