LS 6000 Scintillation System User’s Manual
PN 247971-F
3-20
Counting Samples
Multi-Task
2. From the Multi-Task Menu, highlight “Isotope/DPM Libraries” and press SELECT. The
Main Editing Window displays two prompts: Setup and Review Isotopes and Setup and
Review DPM Quench Curves.
3. Highlight ‘Setup and Review DPM Quench Curves” and press SELECT. The menu
shown in Figure 3.18 is displayed.
4. A new quench curve is manually entered just as if you were manually entering it in the
DPM Setup. Refer to Manual Entry of Quench Curves Section 6 for more information on
entering quench curves manually. You may also edit or delete a stored quench curve.
Refer to Editing Quench Curves Section 6 and Deleting A Quench Curve Section 6,
respectively, for more information.
NOTE You cannot count standards to set up quench curves or print correlation tables in Multi-Task.
5. When you are finished using the DPM Library, press PREVIOUS MENU. The Multi-Task
Menu is displayed.
6. Press PREVIOUS MENU again to return to the Auto Count display.
Figure 3.18 Menu to Edit Quench Curves in Multi-Task.