LS 6000 Scintillation System User’s Manual
PN 247971-F
6-14
Isotope/DPM Libraries
Setting Up A Quench Curve
If you do not wish to count a background quench curve, choose No.
7. Select the type of quench
curve you wish to set up. Highlight “Setup Single Label DPM”,
“Setup Dual Label DPM”, or “Setup Triple Label DPM” and press SELECT. The Main
Editing Window displays prompts to select the isotope(s).
8. The isotopes stored in the Isotope Library are displayed in the Data Entry Window.
Choose the desired isotope. If setting up dual or triple label quench curves, prompts are
presented for each isotope.
If the desired isotope is not displayed, the isotope is not stored in the Isotope Library.
Exit the DPM program by pressing MAIN MENU. Highlight “New Isotope Setup” and
press SELECT. Add the isotope to the library. Refer to Section 6.1 for information on the
Isotope Library.
9. After the appropriate number of isotopes are selected, press the Down Cursor Arrow key.
The menu shown in Figure 6.10 is displayed. The isotope(s) selected is displayed in the
top left side of the menu for your reference.
Figure 6.10 Menu to Setup Quench Curves.