LS 6000 Scintillation System User’s Manual
PN 247971-F
d-2
Isotope Settings and Half-Life
113
Sn-
113
mIn
800 115.1 days-Sn; 99.48 minutes-In
125
I
550 60.25 days
131
I
900 8.04 days
133
Ba
850 10.66 years
137
Cs
850 30.174 years
141
Ce
820 32.55 days
153
Gd
700 241.6 days
195
Au
700 182.9 days
207
Bi
920 38.3 years
210
Pb-
210
Bi-
210
Po
750 22.26 years-Pb: 5.0 13 days-Bi
138.38 days-Po
222
Rn
950 3.824 days
226
Ra
950 1599 years
233
U
820 1.591
x 10
5
years
238
U
750 4.468
x 10
9
years
241
Am
850 432 years
241
Pu
850 14.36 years
Figure d.1 Isotope Settings and Half-Life.
Isotope Channel Half-life
Figure d.2 Universal Half-Life Correction Table.
l/T
FRACTION
REMAINING
t/T
FRACTION
REMAINING
l/T
FRACTION
REMAINING
t/T
FRACTION
REMAINING
0 1,0000 0,52 0,6974 1,54 0,3439 3,80 0,0718
0,01 0,9931 0,54 0,6878 1,56 0,3391 3,85 0,0693
0,02 0,9862 0,56 0,6783 1,58 0,3345 3,90 0,0670
0,03 0,9794 0,58 0,6690 1,60 0,3299 3,95 0,0647
0,04 0,9726 0,60 0,6597 1,62 0,3253 4,00 0,0625
0,03 0,9659 0,62 0,6507 1,64 0,3209 4,10 0,0583
0,06 0,9593 0,64 0,6417 1,66 0,3164 4,20 0,0544