LS 6000 Scintillation System User’s Manual
PN 247971-F
4-13
Setting Up User Programs
Editing A User Program
4
higher energy isotope into the lower energy isotope window essentially constant. It is
selectable.
To change the settings, highlight “Quench/Lum-Ex/2Phase” and press SELECT. The menu
shown in Figure 4.8 is displayed. Choose the desired quench monitor. If H# is chosen, choose
AQC On (Yes) or Off (No). If no other changes are desired, press PREVIOUS MENU to
return to the menu shown in Figure 4.4.
NOTE If a Dual or Triple Label DPM program is selected, the quench monitor is automatically set to H# and
AQC is turned on. This cannot be changed. When Xtal is selected as the scintillator, or when Xtal DPM
program is selected, H# and AQC are not available.
Lum-Ex Correction
Lum-Ex Correction provides a means of distinguishing between counts of actual radioactive
disintegrations within the sample and other light-producing events, such as
chemiluminescence. The % of the total count rate that is luminescence is always printed as %
Lum-Ex on the printout whether it is selected On or Off. When Lum-Ex Correction is
selected On, the luminescence counts in the counting window of each isotope are subtracted.
To change the setting, highlight “Quench/Lum-Ex/2Phase” and press SELECT. The menu
shown in Figure 4.8 is displayed. Highlight “Lum-Ex Correction” (if installed) and choose
NO
(Off) or YES (On). If no other changes are desired, press PREVIOUS MENU to display the
Review/Edit Menu shown in Figure 4.4.
Figure 4.8 Quench/Lum-Ex/2Phase Selection Menu.