LS 6000 Scintillation System User’s Manual
PN 247971-F
b-3
Installation Requirements
Space Requirements
b
lines that are separate from those used to supply power to instruments requiring a constant
power supply.
b.2 Space Requirements
A minimum bench or table space of 132 cm (52 in.) wide and 71 cm (28 in.) deep is required
for the instrument. The surface must be level and flat and capable of supporting 275 kg. (600
lbs.) without bowing or flexing. To allow for optimal air circulation and ease of service, a
recommended space of 152 cm (60 in.) wide and 81 cm (32 in.) deep should be dedicated to
the instrument, an additional 30.5 cm (12 in.) of width (shaded area in Figure b.1) should be
free of permanent structures to allow for installation and service. The area 91 cm (36 in.)
above the bench or table must be clear of shelves or cabinets.
b.3 Environment Requirements
Counting specifications are based on a minimum of i5
0
C (54
0
F) and a maximum of 30
0
C
(86
0
F) ambient temperature, providing calibration has been performed within 5
0
C (9
0
F) on
the operating temperature. Relative humidity must be less than 85% without condensation.
Avoid placing instrument where it is exposed to direct sunlight or severe temperature
fluctuations such as near windows, heaters or air conditioning outlets.
Dimensions:
Width
cm(in.)
Height
cm(in.)
Depth
cm(in.)
Weight
kg(lbs.)
Instrument: 91.5(36) 66(26) 80(31.5) 210(460)
Monitor: 33(13) 34.3(13.5) 35.6(12) 8.2(18)
Printer: 39(15.5) 10(4) 30.5(12) 5(11)
Temperature
Control Accy
35.5(14) 56(22) 73.7(29) 45.4(100)