LS 6000 Scintillation System User’s Manual
PN 247971-F
5-18
Data Calculation
Xtal SL DPM % of Reference
Xtal SL DPM % of Reference (if installed) is available when Xtal is selected as the scintillator.
Single and Dual Label DPM % of Reference (if installed) is available when a liquid scintillator
is selected. Refer to Section 5.7 for information on DPM % of Reference.
Setting Up Xtal SL DPM % of Reference
When editing a User Program, from the Data Calculation Menu (Figure 5.1), choose XTAL SL
DPM%Ref. The prompts displayed on the menu are the same as described for the Xtal DPM
program. Refer to Setting Up Xtal DPM Section 5 for information on these prompts.
Loading Samples for Xtal SL DPM% of Reference
Since the instrument has no way of distinguishing among blanks, standards, references, and
unknowns, a convention has been established for identifying them by the sequence in which
they are loaded. An example loading sequence is shown in Figure 5.16.
Figure 5.15 Typical Printout for DL DPM % of Reference
% Ref
Sample DPM - Blank CDM x 100 x Factor
Reference DPM - Blank DPM
=