LS 6000 Scintillation System User’s Manual
PN 247971-F
5-4
Data Calculation
CPM/Xtal CPM
CPM X Factor = Final Answer Printed Out
To change the factor for any isotope, highlight “Factor” for that isotope and enter the new
factor. The factor may have a maximum of nine digits and decimal point. Values greater than
or less than this range may be entered as an exponential using a 10 character field. For
example, the number 1.0623 X 10- is entered as 1.0623E
23•
Loading Samples for CPM/Xtal CPM
An example loading sequence for a cpm program using blanks and replicates is shown in
Figure 5.2.
When loading samples for cpm, if blanks or replicates are not used, the samples may be
loaded in any manner desired. If blanks are specified in the User Program as described in ID
Section 4, the blanks must be loaded first. Any number of blanks may be used. After the
blanks, an empty space is used, signal-ling the instrument that samples follow. Load the
samples after the empty space.